CS1220 VIMS Features
The CS1220 VIMS ContaminationSensor is a stationary measurement unit for the
continuous monitoring of solid particle contamination in hydraulic and lubrication
systems.
The CS is designed to be used in low- or high-pressure hydraulic and lubrication
circuits and test benches where a small amount of oil (between 30 ml/min and 300
ml/min) is diverted for measurement purposes.
The ContaminationSensor is approved for a maximum operating pressure up to 300
bar (4350 psi) and viscosity of up to 1000 mm²/s.
The particulate contamination is recorded by an optical measurement cell.
The sensor has the following features:
• 4 analog output voltage outputs 0 – 5 V each, 3 for ISO-code and 1 for status
• results are output as a cleanliness code according to:
ISO 4406:1999
• flange installation
The display and the buttons are without function in this model.
CS1000 Restrictions on use
NOTICE
Prohibited operating fluids
The CS will be destroyed.
► Operate the CS1000 only in conjunction with the permitted operating fluids.
► CS 1xx0 is suitable for operation with petroleum or petroleum-based
refinates.
► Observe the maximum operating pressure on the CS 1000 type label.