28 29
Step 3 Applying Control Solution to Test Strip
Shake the bottle before each test.
Remove the cap and squeeze the bottle to
discard the first drop. Then wipe the tip with a
clean tissue or cloth. Dispense a drop of control
solution onto a clean non-absorbent surface.
It helps to squeeze a drop onto the top of the
cap as shown. After the
symbol appears
on the display, apply the solution to the narrow
edge of the test strip until the meter beeps.
Make sure the confirmation window fills
completely.
Note: The meter may switch off if the control solution sample
is not applied within 2 minutes of the
symbol
appearing on the screen. If the meter turns off, remove the
strip, reinsert, and start from step 1.
Step 4 Waiting for the Result
The display segments will rotate
clockwise and a test result will
appear after the meter counts
down from 5 to 1.
The test result with control
solution flag is stored in the
memory but not included in the
averages.
Control Solution Range
Control L :
Control H :
Step 5 Comparing the Result
Compare the result displayed on the
meter to the range printed on the
test strip vial. The result should fall
within the range.
Caution: The range printed on the test strip vial is for the
Control Solution only. It has nothing to do with your blood
glucose level.
Note: The Nipro Premier Control Solution can be purchased
separately. Please contact your authorised Nipro sales
representative.