Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of the range.
Out of range results may occur in following situations:
If results continue to fall outside the range, the test strip and
meter may not be working properly. Do not use your system
and contact Nipro sales representative.
Situations
•Whenthecontrolsolutionbottlewas
not shaken well,
•Whenthemeter,teststrip,orthe
control solution were exposed to high
or low temperatures,
•Whenthefirstdropofthecontrol
solution was not discarded or the tip of
the bottle was not wiped clean,
•Whenthemeterisnotfunctioning
properly.
•Whenthecontrolsolutionispastthe
expiration date printed on the bottle,
•Whenthecontrolsolutionispastits
discard date,
•Whenthecontrolsolutionis
contaminated.
Do This
Repeat the
control solution
test by referring
to the 'Notes' on
page 26.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
Caution:
To avoid infection when drawing a sample, do not use a
lancet more than once, and:
•Donotusealancetthathasbeenusedbyothers.
•Alwaysuseanewsterilelancet.
•Keepthelancingdeviceclean.
Note:
Repeated puncturing at the same sample site may cause pain
or skin calluses (thick hard skin). Choose a different
site each time you test.
Using the Lancing Device
You will need a lancing device in order to collect a blood
sample.
You may use the lancing device that is included in the Nipro
Premier alpha Blood Glucose Monitoring System or any other
medically approved lancing device.
• Thelancingdeviceisforusebyasingleuseronlyandshould
not be shared with anyone.
• Useasoftclothortissuetowipethelancingdevice.If
necessary, a small amount of alcohol on a soft cloth or tissue
may be used.
Adjustable Tip
Lancet Holder
Sliding Barrel
Release Button
Load Confirmation
Window
Ejection System
Lancet