Guidance and Manufacturer’s Declaration IEC 60601-1-2:2014; Edition 4.0
Electromagnetic Immunity
iCare HOME2 tonometer (TA023) is intended for use in a home healthcare environment with
electromagnetic characteristics specified below. The user of the iCare HOME2 tonometer (TA023)
should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment-Guidance
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur
radio
bands between 0,15
MHz and 80 MHz
80 % AM at
1 kHz
3 V
6 V
WARNING!
Portable RF
communications
equipment (including
peripherals such as
antenna cables and
external antennas)
should be used no
closer than 30 cm (12
inches) to any part
of the iCare HOME2
(TA023), including
cables specified by the
manufacturer, to avoid
the degradation of
performance.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
10 V/m
WARNING!
Portable RF
communications
equipment (including
peripherals such as
antenna cables and
external antennas)
should be used no
closer than 30 cm (12
inches) to any part
of the iCare HOME2
(TA023) including
cables specified by the
manufacturer, to avoid
the degradation of
performance.
Interference may
occur in the vicinity
of equipment marked
with the following
symbol: