Guidance and Manufacturer’s Declaration IEC 60601-1-2:2014; Edition 4.0
Electromagnetic Immunity
iCare HOME2 tonometer (TA023) is intended for use in a home healthcare environment with
electromagnetic characteristics specified below. The user of the iCare HOME2 tonometer (TA023)
should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment-Guidance
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
380 - 390 MHz
27 V/m; PM 50%;
18 Hz
430 - 470 MHz
28 V/m; (FM ±5
kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%;
217 Hz
800 - 960 MHz
28 V/m; PM 50%;
18 Hz
1700 - 1990 MHz
28 V/m; PM 50%;
217 Hz
2400 - 2570 MHz
28 V/m; PM 50%;
217 Hz
5100 - 5800 MHz
9 V/m; PM 50%;
217 Hz
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
WARNING!
Portable RF
communications
equipment (including
peripherals such as
antenna cables and
external antennas)
should be used no
closer than 30 cm (12
inches) to any part
of the iCare HOME2
(TA023) including
cables specified by the
manufacturer, to avoid
the degradation of
performance.
Interference may
occur in the vicinity
of equipment marked
with the following
symbol: