EasyManua.ls Logo

idronaut Ocean Seven 310 - Page 103

idronaut Ocean Seven 310
139 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPENDIX C OPTIONAL SENSORS
IDRONAUT Brugherio (MB) OCEAN SEVEN 310 CTD 01-2019
92
Measuring scale 2, Full scale [ 500 FTU]
Do you already know 'SLOPE & OFFSET' coefficients ?:No type [Y]es or [N]o < y
Offset:0
Slope:0 enter new value< 1
Measuring scale 3, Full scale [ 125 FTU]
Do you already know 'SLOPE & OFFSET' coefficients ?:No type [Y]es or [N]o < y
Offset:0
Slope:0 enter new value< 1
Measuring scale 4, Full scale [ 25 FTU]
Do you already know 'SLOPE & OFFSET' coefficients ?:No type [Y]es or [N]o < y
Offset:0
Slope:0 enter new value< 1
Update calibration parameters
Turbidity meter full scale:(1)>750 FTU,(2)500 FTU,(3)125 FTU,(4)25 FTU
Measuring range [1..4]:1
The STM sensor calibration is complete and the CTD shows the list of sensors. The procedure
foresees that the slope and offset coefficients for each measuring scale is known. It is possible to
calculate offset and slope starting from a table of values by answering NO at the question about the
offset and slope. Contact IDRONAUT to receive detailed instructions.

Table of Contents

Related product manuals