Offset:0
Slope:0 enter new value< 1
Measuring scale 2, Full scale [ 50]
Do you already know 'SLOPE & OFFSET' coefficients ?:No type [Y]es or [N]o < y
Offset:0
Slope:0 enter new value< 1
Measuring scale 3, Full scale [ 15]
Do you already know 'SLOPE & OFFSET' coefficients ?:No type [Y]es or [N]o < y
Offset:0
Slope:0 enter new value< 1
Measuring scale 4, Full scale [ 5]
Do you already know 'SLOPE & OFFSET' coefficients ?:No type [Y]es or [N]o < y
Offset:0
Slope:0 enter new value< 1
Update calibration parameters
Fluorometer full scale:(1)>150,(2)50,(3)15,(4)5
Measuring range [1..4]:1
The STF sensor calibration is complete and the CTD shows the list of sensors. The procedure foresees
that the slope and offset coefficients for each measuring scale is known. It is possible to calculate
offset and slope starting from a table of values by answering NO at the question about the offset and
slope. Contact IDRONAUT to receive detailed instructions.