APPENDIX C OPTIONAL SENSORS
IDRONAUT Brugherio (MB) OCEAN SEVEN 310 CTD 01-2019
96
The PAR sensor is calibrated by entering an offset and slope for each measuring range. The
procedure can be run starting from the CASE command available under the CALB menu.
PAR SENSOR
Do you already know 'SLOPE & OFFSET' coefficients ?:No type [Y]es or [N]o < y
Offset:0
Slope:0 enter new value< 1
The STF sensor calibration is complete and the CTD shows the list of sensors