APPENDIX C OPTIONAL SENSORS
IDRONAUT Brugherio (MB) OCEAN SEVEN 310 CTD 01-2019
104
C10 CYCLOPS Fluorometer
The TURNER DESIGNS Cyclops submersible sensor is an accurate single-channel detector that can
be used for many different applications. It is designed for integration into multi-parameter systems
from which it receives the power and delivers a voltage output proportional to the concentration of
the fluorophore, particle or compound of interest.
The Cyclops voltage output can be correlated to concentration values by calibrating with a standard
of known concentration.
The Cyclops output can be scaled x1, x10, x100.
C10.1 CONFIGURATION
The following parameters are requested to properly set up the Cyclops 7 Fluorometer. The sensor
setup sequence can be run using the CNAP command available under the CONF menu
Logical code [0..255]:201 or 202
Description :Chl-a
Significant digits [1..13]:6
Digits after the dot [0..9]:2
Do you want to store it ?:Yes
Do you want to show data ?:Yes
Data processing method
0)Counts,1)mV,2)Custom&UNESCO,3)Polynomial,4)a+bx
Method [0..4]:3
Mux[0..7,240=A.Board,255=NU]:1
Compensate thermal drift ?:No
Cyclops7F gain 1)x1,2)x10,3)x100
C10.2 CALIBRATION
The Cyclops sensor is calibrated by entering the calibration coefficients for the selected measuring
scalee. The procedure can be run starting from the CASE command available under the CALB menu.
TURNERDESIGN CYCLOPS 7 - Calibration
Select CYCLOPS7 gain 1)x1,2)x10,3)x100
Select the working measuring scale
Number of coefficients: 2
Enter the number of calibration coefficients and then the value for each ones.
Coeff. a:
Coeff b: