SECTION SIX SERVICE, DIAGNOSTIC AND CONFIGURATION
IDRONAUT Brugherio (MB) OCEAN SEVEN 310 CTD 01-2019
68
The probe sampling rate
Sampling rate [1..28]Hz:28
6.1.3.6 Operating parameters
Through this command, it is possible to customize some aspects of the probe operation. The
customizable parameters are:
OCEAN SEVEN 310 - ID:0317003 {USR}(1.1_00 06/17)Oct 01 12:26:28.06 2016
--- Digital setup ---Config[6456]Status[488]
Warmup timeout [s]:5
The operator can customize the time-out used by the probe at the wake-up to wait for the sensor
stabilization. The default value is 5. In case unattended profiles or monitoring must be performed and
the probe is equipped with chemical sensors, we suggest a warm-up of 10s. If the GO1018 rosette is
interfaced by the probe we suggest a timeout of 30s (unattended operations).
Realtime data at startup ?:No
Is it possible to configure the probe to automatically show at startup data acquired in real time
Number of lines between header :22
This parameter allows the operator to configure the number of data lines that will be shown between
two consecutive data header lines.
Max. number of char. per line :160
This parameter allows the operator to configure the number of characters per line. The data acquisition
functions use this parameter to break data message below this limit.
6.1.3.7 Acquired sensors
Through this command, it is possible to customize the type and number of the interfaced sensors. Once
invoked, the list of the acquired sensors appear on the PC screen as follows:
OCEAN SEVEN 310 - ID:0317003 {USR}(1.1_00 06/17)Oct 01 12:32:03.01 2016
Probe parameter list
Idx Parameter Mux Type Code Cal.Rec
01 Press 240 24bit-Ana 010 000 000
02 Temp 240 24bit-Ana 009 001 001
03 Cond 240 24bit-Ana 008 002 002
04 Sal 255 -------------- 004 ---<<--P{0}T{1}C{2}
05 O2Sat% 240 24bit-Ana 000 006 005<<--T{1}
06 O2ppm 255 -------------- 005 ---<<--T{1}O2{6}S{0}
07 pH 240 24bit-Ana 001 007 007<<--T{1}
08 Eh 240 24bit-Ana 002 008 004
CMD:(A)dd,(D)elete,(M)odify,(Q)uit