The list of the acquired sensors ends with the commands available for the customization list.
6.1.3.8 Available commands
Add/ Modify The add/modify command allows the operator to add a new sensor or to
modify one sensor already configured. When you select the modify function,
the following message appears:
Modify the parameter: 0
The parameter/sensor to be modified or added must be selected by means of
the index indicated in the first column of the above list. The add function will
insert a new parameter at the selected position moving the remaining
parameters in the list, one position below the insertion point. In both cases,
the following answers appear on the PC screen:
Logical code :12
Description :Tr%
N. of digits to show :0
N. of digits after the dot :0
Do you want to store it ? :No
Do you want to show data ? :No
Dataprocessing method 0)Counts,1)mV,2)Custom&UNESCO,3)Polynomial,4)a+bx
Data processing method :4
Mux. channel [255=none] :0
Mux. delay [0..255ms ]:0
See below for a complete description of the above parameters.
Initialize Choosing the “Initialize” command will cause the acquired parameter to be
preset to a default condition. Anyhow, please be aware that calibrations are
lost if this command is invoked.
Delete Choosing the “Delete” command will cause the following message to appear:
Delete the parameter: 0
By means of the index, the operator must select the parameter to delete.
Quit Terminates the configuration function. When the “Quit” command is selected,
the “Configuration Menu” is shown.
Typical parameters that can be configured for each sensor in the above list are:
Logical code Assigns a logical code to the sensor, from the list of pre-defined
sensors/parameters (see below).
Description A seven-character acronym that is used by the probe to identify data acquired
from this sensor.
No. of digits to show Number of digits used by the probe to numerically show the acquired data
for this sensor.