EasyManua.ls Logo

IFM Electronic LR3000 - 6�4�3 Installation in Open Tanks

IFM Electronic LR3000
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
UK
Arrange for a bore hole in the tank lid� It must have a minimum diameter (d) to
enable sufficient transfer of the measured signal to the probe (fig� 6-10)� The
diameter depends on the wall thickness of the tank lid:
Wall thickness [mm] 1���5 5���8 8���11
Bore hole diameter [mm] 35 45 55
Install the flange plate with the flat surface showing to the tank and fix it with
appropriate screws�
A seal (B in fig� 6-11) can be inserted between flange plate and tank� Some
flange plates are supplied with a seal�
Ensure cleanness and evenness of the sealing areas, especially if the tank is
under pressure� Tighten the fixing screws sufficiently
Screw the unit into the flange plate using the process connection and tighten
firmly
Make sure that the supplied sensor seal (A in fig� 6-11) is correctly positioned�
6.4.3 Installation in open tanks
For installation in open tanks, use a metal fixture to install the unit� It serves
as a launching plate (R); minimum size: 150 x 150 mm for a square fixture,
150 mm diameter for a circular fixture�
If possible, mount the unit in the middle of the fixture� The distance D2 must not
be below 40 mm, higher for probe lengths > 70 cm and in case of heavy soiling
(→ 6.1.1):
D2
150 mm
R
The lower edge of the process connection should be flush with the installation
environment (see fig� 6-7)�
Avoid non-flush installation (see fig� 6-8)�

Table of Contents

Related product manuals