EasyManua.ls Logo

IFM Electronic SM6000 - 9.4 Settings for temperature monitoring

IFM Electronic SM6000
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
9.3.2 Settings for quantity monitoring using the preset meter
Select [OU1] and set [ImP]�
Select [ImPS] and set the volumetric flow quantity at which output 1
switches(→9.7).
Select [ImPR] and set [no]: pulse repetition is not active� The output
switches ON if the value set in [ImPS] is reached� It remains switched
until the meter is reset�
9.3.3 Settings for meter reset controlled by the program
Select [rTO] and continue with a) or b)�
a) Reset the meter manually: press [Set] until [rES�T] is displayed�
Press [Mode/Enter] briefly
b) Enter the value for time-controlled reset: Press [Set] until the re-
quested value is displayed (intervals from 1 hour to 8 weeks)�
Press [Mode/Enter] briefly
9.3.4 Switch off the meter reset
Select [rTO] and set [OFF]� The meter is only reset after overflow
(= factory setting)�
9.3.5 Settings for meter reset by an external signal
Select [OU2] and then [InD]�
Select [Din2] and set the reset signal:
- [Hi] = reset for high signal,
- [Lo] = reset for low signal,
- [+EDG] = reset for rising edge,
- [-EDG] = reset for falling edge�
9.4 Settings for temperature monitoring
9.4.1 Settings for limit value monitoring with OUT2
Select [SEL2] and set [TEMP]�
Select [OU2] and set the switching function:
- [Hno] = hysteresis function/normally open,
- [Hnc] = hysteresis function/normally closed,
- [Fno] = window function/normally open,
- [Fnc] = window function/normally closed�
Select [SP2] and set the value at which the output switches�
Select [rP2] and set the value at which the output switches back�

Table of Contents

Related product manuals