EasyManua.ls Logo

IFM Electronic SM6000 - 9.6 Service functions

IFM Electronic SM6000
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
9.5.6 Setting the error behaviour of OUT1 / OUT2
Select [FOU1] and determine the value:
- [On] = output 1 switches ON in case of an error
- [OFF] = output 1 switches OFF in case of an error
- [OU] = output 1 switches irrespective of the error as defined with the
parameters�
Select [FOU2] and determine the value:
- [On] = output 2 switches ON in case of an error, the analogue signal
goes to the upper end stop value�
- [OFF] = output 2 switches OFF in case of an error, the analogue signal
goes to the lower end stop value�
- [OU] = output 2 switches irrespective of the error as defined with the
parameters� The analogue signal corresponds to the measured value�
9.6 Service functions
9.6.1 Reading the min./max. values for volumetric flow
Select [HI] or [LO] and press [Set] briefly
[HI] = maximum value, [LO] = minimum value�
Delete memory:
Select [HI] or [LO]�
Press [Set] and keep it pressed until [----] is displayed�
Press [Mode/Enter] briefly
It makes sense to delete the memories as soon as the unit works under
normal operating conditions for the first time�
9.6.2 Reset all parameters to the factory setting
Select [rES], then press [Set] and keep it pressed until [----] is dis-
played�
Press [Mode/Enter] briefly
Thefactorysettingislistedattheendoftheinstructions(→13Factory
setting)�
It makes sense to write your own settings in this table before executing the
function�

Table of Contents

Related product manuals