Important Information:
If the absorbance value of the sample is not in the linear range the following “Warning messages” will appear and
“Instruction” will be displayed in the top left corner of the result screen.
Concentration too high.
Do you want to change the lid factor?
Please change to lid 10 and press sample.
(automatic change of lid factor lid 5 to lid 10 in the
software for calculation)
Concentration too high.
Do you want to change the lid factor?
Please change to lid 50 and press sample.
(automatic change of lid factor lid 10 to lid 50 in the
software for calculation)
Concentration too low.
Do you want to change the lid factor?
Please change to lid 10, apply a minimum of 1µl of
sample and press sample. (automatic change of lid
factor lid 50 to lid10 in the software for calculation)
Concentration too high.
Dilute sample or change to lid 100.
Concentration too low.
Do you want to change the lid factor?
Please change to lid 50 and press sample.
(automatic change of lid factor lid 100 to lid 50 in
the software for calculation)
Concentration too high.
Dilute sample or change to lid 250.
Concentration too low.
Do you want to change the lid factor?
Please change to lid 100 and press sample.
(automatic change of lid factor lid 250 to lid 100 in
the software for calculation)
Concentration too high.
Dilute sample.
Sample concentration is too low
Sample concentration is too low (or change to lid 5 if
available)
Physical dilution of the sample is necessary (or change
to lid 100 if available)
Physical dilution of the sample is necessary (or change
to lid 250 if available)
Physical dilution of the sample is necessary
*Some of the lids are only optional available. Lid delivery content for P 300 is Lid 50, for P 330 Lid 10 and Lid 50 and
for P360 is Lid 10, Lid 50 und Lid 250.