NanoPhotometer
®
P-Class User Manual
Version 2.1 Page 69 / 70
12.3 Protein quantification
For determination of protein concentration in solution the absorbance at wavelength 280 nm is used. The function
describing the concentration to absorbance relation is a modification of the Lambert-Beer equation.
c
prot.
= Abs. 280 * A280 factor * lid factor * dilution factor
With background correction:
c
prot.
= (Abs. 280 Abs. 320) * A280 factor * lid factor * dilution factor
c
prot
protein concentration (mg/ml)
Abs. 280 absorbance (AU) of proteins
A280 factor Default setting is BSA; molecular weight
prot.
/ molar extinction coefficient (M
-1
*cm
-1
)
prot.
or
1 / extinction coefficient (l/g*cm)
lid factor virtual dilution factor 5, 10, 50, 100 and 250; dependent on the used dilution lid.
A280 factors pre-programmed:
BSA
Serum
albumin
(mouse)
Serum albumin
(human)
IgG
Lysozyme
A280 factor
1.45
1.57
1.76
0.71
0.38
Molecular Weight [g/mol]
69,323.4
68,692.5
69,365.7
150,000
14,300
Molar extinction coefficient [M
-1
*cm
-1
]
47,790
43,780
39,310
210,000
37,500
12.4 Protein fluorescent dye incorporation
To determine the protein concentration and the dye concentration after labelling a modification of the Lambert-Beer
equation is used. Background correction is always calculated; possibility to switch the dye correction on and off.
Calculation of labelled protein concentration
c
prot
= (Abs. 280 Abs. 320) * A280 factor * lid factor * dilution factor
With dye correction:
c
prot
= [Abs. 280 Abs.320 - CF
dye
* (Abs.
max, dye
Abs. 320)] * A280 factor * lid factor * dilution factor
c
prot
protein concentration (mg/ml)
Abs. 280 absorbance (AU) of proteins
CF
dye
dye-dependent correction factor at 280 nm (to be delivered from dye-supplier)
Abs.
max, dye
absorbance at absorption maximum of the dye (AU)
A280 factor molecular weight
prot.
/ molar extinction coefficient (M
-1
*cm
-1
)
prot.
or
1 / extinction coefficient (l/g*cm)
lid factor virtual dilution factor 5, 10, 50, 100 and 250; dependent on the used dilution lid.
Calculation of fluorescence dye concentration [pmol/µl]
c
dye
= (Abs.
max, dye
Abs. 320) * lid factor * dilution factor /
dye
* 10
-6
)
c
dye
dye concentration (pmol/µl)
Abs.
max, dye
absorbance at absorption maximum of the dye (AU)
lid factor virtual dilution factor 5, 10, 50, 100 and 250; dependent on the used dilution lid.
ε dye dye-dependent molar extinction coefficient (M
-1
* cm
-1
)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen P 330 and is the answer not in the manual?

Implen P 330 Specifications

General IconGeneral
BrandImplen
ModelP 330
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals