EasyManuals Logo

Implen P 330 User Manual

Default Icon
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
NanoPhotometer
®
P-Class User Manual
Version 2.1 Page 16 / 70
4.1.3 Analysis of Oligonucleotides
The procedure is as follows:
Parameter Screen
NanoVolume Applications
Cuvette Applications
Parameter Screen
Step 1 Press 1 for NanoVolume OR 2 for Cuvette folder.
Step 2 Press 1 to select Nucleic Acids folder.
Step 3 Press 4 to select Oligo mode.
Step 4 Using the NanoVolume Applications select the Lid Factor as
described in the “Average Detection Range Sheet” and
under 3.2.
Step 5 Enter the Dilution Factor using the keypad numbers. Range
1.00 to 9,999. Use the C button to backspace and clear the
last digit entered OR press Menu/Options to enter the
dilution factor screen. Enter the volume of the sample
using the keypad numbers. Range 0.01 to 9,999. Enter the
volume of the diluent using the keypad numbers. Range
0.01 to 9,999. Press OK to calculate the dilution factor
and return to the Parameters screen OR press Cancel to
cancel the selections and return to the Parameters screen.
Step 6 Background correction at 320 nm is recommended to be
switched on.
Step 7 Select the Units of measurement using the left and right
arrows. Options: μg/ml, ng/μl, μg/μl and pmol/μl.
Step 8 Enter the Factor using the keypad numbers. Default value
is 33, range is 0.01 to 9,999.
Step 9 If pmol/μl is selected there are two options to set the factor
1. A selection table denoting the ratios of the 4 bases
according to the oligo sequence. Enter the proportions of
bases present using the keypad numbers and up and down
arrows to move between boxes. Default is 10 for each,
range is 0 to 9,999.
2. Enter the known extinction factor of the oligo used:
factor range 0.01 to 9,999 for ratio = [1 / extinction
coefficient *10
-6
].
Step 10 Press OK to enter the Results screen OR Cancel to
return to the Nucleic Acids folder.
Results Screen
Results Screen
Step 11 Apply/insert the reference sample. Press Blank Key. This
will be used for all subsequent samples until changed.
Step 12 Apply/insert sample and press Sample . This measures
at the selected wavelengths and displays the results. The
sample concentration and the ratio of A260/A280 and
A260/A230 are calculated (corrected by the background
wavelength value if selected).
Step 13 If the absorbance value of the sample is not in the linear
range a “Warning message will pop up and “Instruction”
will be displayed in the top left corner of the result screen.
Please refer to 3.2 Software instructions/important
information on page 11 for further information.
Step 14 Repeat for all samples.
Step 15 Press Menu/Options to display available Options which are
described on page 8.
Step 16 Press Escape and confirm with Yes to return to the
Nucleic Acids folder.
To change parameters, print or save methods press the Menu/Options button. The options menu will be opened. For
further explanation please see 2.3 Keypad and display on page 6 (P 300) and 7 (P330 / P 360).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen P 330 and is the answer not in the manual?

Implen P 330 Specifications

General IconGeneral
BrandImplen
ModelP 330
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals