IMT Analytics FlowAnalyser PF-300 PRO - Power Supply and Battery Operation

IMT Analytics FlowAnalyser PF-300 PRO
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
14
4
IMT Analytics AG
Gas standard Temperature Pressure Relative
humidity
API Standard Conditions 15/1013
15
°C (60 °F)
1013.25 mbar
(14.7 psia)
0 %
Cummings Standard 25/991
25
°C (77 °F)
991 mbar
(500 ft altitude)
0 %
20 °C/1013 mbar 20/1013
20
°C (68 °F)
1013.25 mbar
(760 mmHg)
0 %
Normal Temperature and
Pressure
NTPD
20.0
°C (68 °F)
1013.25 mbar
(760mmHg)
0 %
Normal Temperature and
Pressure, Saturated
NTPS
20.0
°C (68 °F)
1013.25 mbar
(760mmHg)
100 %
Ambient Pressure at 25 °C AP25
25.0
°C (77 °F)
Current Ambient
pressure
Current gas
humidity
In this User manual, the unit L/min is based on ambient conditions of
0 °C and 1013 mbar (DIN 1343).
Please refer to Appendix B: Measurement parameters and units.
Here, you will also find the conversion factors for the units of meas-
urement.
Input voltage of the power supply unit: 100 240 VAC, 50 60 Hz, 2.0 A
Output voltage of the power supply unit: 24 VDC, 2.7 A
PF-300 PRO input voltage: 24 VDC, 2.5 A
Power supply protection class: Class I
Only use the power supply unit and cable provided!
Operating time in battery operation: 16 hours
Operating time in battery operation
with MultiGasAnalyser: 13 hours
Charging the battery
A complete charging process takes 8 h. The service life of the battery is extended if
the battery is charged completely only after a prompt by the device.
Thedevice indicates visually and audiblywhen the battery must be
charged. Please do not store the battery in depleted state.
Caution:depletion can damage the battery beyond repair!
4.3 Power supply
4.4 Battery operation