Specifications
9
4
IMT Analytics AG
4.1.1 Measuring instrument values
1
Flow Range
Accuracy
-300 – 300
L/min***
± 1.65 %* or ± 0.04 L/min
(for 10 – 40° C)**
Ultra-low flow Range
Accuracy
-1 – 1 L/min***
± 1.65 %* or ± 0.01 L/min
(for 10 – 40° C)**
Volume Range
Accuracy
0 – 1000 L
± 1.75 %* or ± 0.10 mL
Pressure (at flow) Range
Accuracy
-50 – 160 mbar
± 0.5 %* or ± 0.1 mbar**
High differential pressure Range
Accuracy
-250 – 250 mbar
± 0.5 %* or ± 0.1 mbar**
Low differential pressure Range
Accuracy
-10 – 10 mbar
± 1 %* or ± 0.01 mbar**
High pressure and
vacuum
Range
Accuracy
-1 – 10 bar
± 1 %* or ± 7 mbar**
Ambient pressure Range
Accuracy
500 – 1240 mbar
± 1 %* or ± 5 mbar**
Oxygen Range
Accuracy
0 – 100 vol %
± 1 vol %**
Temperature Range
Accuracy
0 – 50
°C
± 1.75 %* or ± 0.5 °C**
Humidity Range
Accuracy
0 – 100 % RH***
± 3 % RH** from 10 % RH to 80 % RH
± 5 % RH** for <10˚% RH and > 80 % RH
1
Standard litres per minute (converted to STP conditions of 21.1°C and 1013 mbar)
* Tolerance related to the measured value
** Ablsoute tolerance, with steady air flow
*** Non-condensing