Servicing and maintenance
10
IMT Analytics AG
Careful servicing in compliance with the instructions is essential for ensuring that
FlowAnalyser PROoperates safely andefficiently. Only components recommended
by the manufacturer may be used.
It isabsolutely essential tocomply with the guidelines andservicing
instructions issued by thevarious manufacturers.
The servicing operations listed below may only be performed by per-
sons whoare familiar with FlowAnalyser PRO. All further repair work
may only be performed by authorised trained professionals. Please
also observe the information issued by the various manufacturers.
Toensure that your deviceoperates with precision and reliabilityfor as long as possi-
ble,it is essential to perform the following servicing routines regularly while the device
is switched off:
During operation
Use of the filter included
After each use
The outer surface of the device enclosure shall withstand wipe disinfection with a
cloth moistened with ethanol or isopropyl alcohol.
Every four weeks
Visually check the protection filter for dust or contamination. Replace the filter if con-
tamination is visible.
Every 12 months:
Factory calibration to ensure reliable measurement. To have FlowAnalyser PRO
calibrated at the manufacturer's, IMT Analytics AG, visit the website
www.imtanalytics.com/easycal