Start-up
21
5
IMT Analytics AG
5.3.6 Differential pressure
The differential pressure connectors can be used for differential pressure measure-
ments.
Measuring range: -250 – 250 mbar
Accuracy: ± 0.5 % of reading or ± 0.1 mbar
5.3.7 High pressure and vacuum
The high-pressure port can be used for measuring pressures from -1 to 10 bar. If a
DISS-O
2
connector is required for the port, an appropriate adapter can be ordered.
Measuring range: -1 –10 bar
Accuracy: ± 1 % of reading or ± 7 mbar