pressure) to overpressure (casting pressure) inside the chamber.
The generator stops heating. On the display “Time” a timer counts
the seconds after casting.
> After a sufficient waiting time, unlock the handle and tilt the unit
back to horizontal position. The casting pressure inside the cruci-
ble chamber is released through “Gas Out” to ambient pressure.
> Open the casting chamber and remove flask and flask holder.
Attention!
You have to activate the melting pressure by pressing “Start”-
button after the washing cycles.
7.4 Error diagnosis
There two types of trouble.
Error and
Warnings.
If an error occur the heating will be switched off and you’ll see error
code in display.
With a light fault a warning appear in display. You’ll see the warning
code in display.
7.5 Troubleshooting
Only an expert may open the system.
Machine can’t
be switched on.
Cooling water supply not switched on („E010”).
Protective gas supply not switched on („E013“).
Thermocouple not put in or faulty („0FbE“).
Generator overheated (too hot, „E021“).
Another error („Exxx“).
Exhibition mode activated (P.155 = 0?).
Temperature
indication not
right
Wrong thermocouple programmed, see software doc-
umentation.
Thermocouple faulty.
Because of that the generator can stop heating!
Set-value of temperature too low.
Additional error messages see software-documentation.
7.6 Service
If you need technical support from company INDUTHERM
Erwärmungsanlagen GmbH, we’d like to have following information
with first contact:
Service No. M19999 DM on the left front side of machine or
Service No. from nameplate at backside of machine.