tina56e1-f     (2015-09)  59 
  If the pressure value 4.99×10
-5
 mbar is output at the 
measurement value output or on the LCD display, the 
adjustment has been successful. Otherwise, repeat the 
adjustment procedure. 
 
 
 
6.2  Adjusting the Atmospheric Pressure Sensor 
The ambient pressure of the gauge is measured by a separate 
atmospheric pressure sensor built into the electronics unit of the 
gauge. 
The atmospheric pressure sensor can be calibrated against the 
diaphragm capacitive sensor in the gauge. The gauge 
electronics compares the output signals of the two sensors and 
carries out the necessary adjustments to the atmospheric 
pressure sensor signal. 
  The adjustment of the atmospheric pressure sensor can 
only be carried out via 
•  the diagnostic port (→  [3]) 
•  the DeviceNet, Profibus, EtherCAT and RS485 
interface (→  [3], [4], [5], [6]). 
 
 
 
6.3 Troubleshooting 
In the event of a fault or a complete failure of the output signal, 
the gauge can easily be checked. 
 
 
Required tools / material 
• Voltmeter / ohmmeter 
•  Allen wrench, AF 2 
•  Spare sensor (if the sensor is faulty)