4.6 DeviceNet Operation 
 
 
Caution
 
Caution: data transmission errors 
The attempt to operate the DeviceNet gauge with 
the RS232C interface causes data transmission 
errors. 
This DeviceNet gauge must not be operated with 
the RS232C interface. 
 
 
Before the gauge is put into operation, it has to be configured for 
the DeviceNet. A configuration tool and the device specific EDS 
(Electronic Data Sheet) file are required for this purpose. This 
software can be downloaded from our website. 
 
 
Node Address Setting  (default 63
dec
) 
 
 
Set the node address (0 … 63
dec
) via the 
<ADDRESS>, <MSD>, and <LSD> switches 
(default 63
dec
). The node address is polled by 
the firmware when the gauge is switched on. 
If the setting deviates from the stored value, 
the new value is taken over into the NVRAM. 
If a setting higher than 63 is made, the pre-
vious node address setting remains valid. 
If the <MSD> switch is in the <P> position, 
the node address is programmable via the 
DeviceNet (→  [4]). 
Example:  Node address = 63: 
P
6
MSD
 
2
4
LSD