12 IG9972BEN (2017-09) Plug-In Boards.om
5 Adjustment
Pirani measurement circuits are factory-adjusted to the standard gauge. Due to
manufacturing tolerances, contamination of the gauges, and different cable
lengths, deviations are inevitable.
Cold cathode measurement circuits are factory-adjusted and require no readjust-
ment.
Two trimmer potentiometers are available for compen-
sating the gauge tolerances, gauge contamination, or
different cable lengths, within certain limits.
The adjustment should only be performed
after the equipment has attained operating
temperature (10 minutes).
With gauge connected, lower the vacuum chamber pressure to <1×10
-4
hPa
Select the measurement circuit to be adjusted (»sensor« mode)
With the <HIGH VAC> potentiometer adjust the display to 8.0×10
-4
hPa
Turn the potentiometer clockwise by 90°
«ur 10
-4
» should now be displayed ( [1]).
Expose the gauge to atmospheric pressure (vent the vacuum chamber)
Turn the <ATM> potentiometer to obtain a reading of 1.0×10
3
hPa
Decrease the pressure to <1×10
-4
hPa
Check the high vacuum reading and readjust, if necessary.
5.1 Adjusting the Pirani
Measurement Circuit
Adjustment at High Vacuum
Adjustment at Atmospheric
Pressure