IG9972BEN     (2017-09)     Plug-In Boards.om  9 
 
  Additional protective measures must be taken if certain processes in the 
vacuum system (e.g. flashovers) can cause hazardous voltages on the 
gauge terminals. 
 
  Although the gauge cables are screened, they should not be routed in 
parallel to lines producing strong electrical noise. 
 
Connect the gauge to the <IKR> connector. Coaxial cables normally suffice. The 
following diagram indicates the conditions under which a triaxial cable is required. 
 
 
 
  The maximum length of 100 m for coaxial cables is specified by 
EN 61010. Greater lengths are not admissible without additional protec-
tive measures. 
If the gauge is not grounded via the vacuum chamber, it must be 
grounded separately. 
  If the cable length is >100 m (only admissible with triaxial cable), the 
connectors must be protected against unintentional separation and 
contact of the center conductor. The cable must only be plugged in or 
detached while the unit is switched off for at least 15 seconds. 
4.3  Connecting the Cold 
Cathode Gauge