tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 37
Value
SP1-H I The upper switching threshold (Setpoint high)
defines the pressure at which the switching
function is deactivated when the pressure is
rising.
e.g.: SP1-H 1500 I gauge dependent.
If another gauge type is connected, the
VGC50x automatically adjusts the threshold
if required.
SPx-H min. SPx-H max.
PSGxxx I
= SPx-L min.
1×10
3
PCGxxx I
1.5×10
3
PEG100/MAGxxx I
1×10
-2
MPGxxx I
1×10
3
BPG400 I
1×10
3
BPG402 I
1×10
3
HPG400 I
1×10
3
BCG450 I
1.5×10
3
CDGxxx I
F.S.
CDGxxxD I
F.S.
all values in mbar, GAS=nitrogen
The minimum hysteresis between the upper and lower switching
threshold amounts to at least 10% of the lower threshold (logarithmic
gauges) or 1% of the full scale value (linear gauges). This prevents
unstable states.
Limits of the upper switching
thresholds