tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 91
The group is only available when a USB memory stick formatted for the
the FAT file system (FAT32) is plugged in. Use a max. 32 GB memory
stick.
Transmit: DAT [,yyyy-mm-dd] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: yyyy-mm-dd <CR><LF>
Description
yyyy-mm-dd Current date in the format yyyy-mm-dd
Transmit: LCM [,a,b,c,ddddddd] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c,ddddddd <CR><LF>
Description
a Data logger command, a =
0 –> Stop / data logging stopped
1 –> Start / data logging started
2 –> Clear / deletion of measurement data file
(ending CSV) from USB memory stick
b Data logging interval, b =
0 –> Logging interval 1/s
1 –> Logging interval 1/10 s
2 –> Logging interval 1/30 s
3 –> Logging interval 1/60 s
4 –> Logging interval in the event of measurement
value changes ≥1%
5 –> Logging interval in the event of measurement
value changes ≥5%
c Decimal separator, c =
0 –> , (decimal comma)
1 –> . (decimal point)
ddddddd File name (max. 7 digits)
Transmit: TIM [,hh:mm] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: hh:mm <CR><LF>
Description
hh:mm Current time in the format hh:mm [24 h]
5.9 Data Logger Parameters
DAT - Date
LCM - Start / stop data logger
TIM - Time