EasyManua.ls Logo

Inficon VGC501 - Example

Inficon VGC501
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
98 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
"Transmit (T)" and "Receive (R)" are related to host.
T: TID <CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: PSG <CR> <LF>
Request for gauge identification
Positive acknowledgement
Request for data transmission
Gauge identification
T: SP1 <CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: 1,1.0000E-09,9.0000E-07 <CR> <LF>
Request for parameters of switching function 1
(setpoint 1)
Positive acknowledgement
Request for data transmission
Thresholds
T: SP1,1,6.80E-3,9.80E-3 <CR> [<LF>]
R: <ACK> <CR> <LF>
Modification of parameters of switching
function 1 (setpoint 1)
Positive acknowledgement
T: FOL,2 <CR> [<LF>]
R: <NAK> <CR> <LF>
T: <ENQ>
R: 0001 <CR> <LF>
T: FIL,2 <CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: 2 <CR> <LF>
Modification of filter time constant (syntax error)
Negative acknowledgement
Request for data transmission
ERROR word
Modification of filter time constant
Positive acknowledgement
Request for data transmission
Filter time constants
T: PR1 <CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: 0,8.3400E-03 <CR> <LF>
T: <ENQ>
R: 1,8.0000E-04 <CR> <LF>
Request for measurement data
Positive acknowledgement
Request for data transmission
Status and pressure
Request for data transmission
Status and pressure
5.13 Example

Related product manuals