EasyManua.ls Logo

Inficon VGC501 - Switching Function Parameters

Inficon VGC501
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
Transmit: SPS <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c,d,e,f <CR><LF>
Description
a Status switching function 1, a =
0 –> Off
1 –> On
b Status switching function 2
c Status switching function 3
d Status switching function 4
e Status switching function 5
f Status switching function 6
Transmit: SPx [,a,x.xxxxEsxx,y.yyyyEsyy] <CR>[<LF>]
Description
x Switching function, x =
1 –> Switching function 1
2 –> Switching function 2
3 –> Switching function 3
4 –> Switching function 4
5 –> Switching function 5
6 –> Switching function 6
a Switching function assignment, a =
0 –> Turned off
1 –> Turned on
2 –> Measurement channel 1
3 –> Measurement channel 2
4 –> Measurement channel 3
x.xxxxEsxx Lower threshold
1)
[in current pressure unit]
(default = depending on gauge) (s = sign)
y.yyyyEsyy Upper threshold
1)
[in current pressure unit]
(default = depending on gauge) (s = sign)
1)
Values can be entered in any format.
They are internally converted into the floating point format.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,x.xxxxEsxx,y.yyyyEsyy <CR><LF>
Description
a Switching function assignment
x.xxxxEsxx Lower threshold [in current pressure unit]
(s = sign)
y.yyyyEsyy Upper threshold [in current pressure unit]
(s = sign)
5.5 Switching Function
Parameters
SPS - Switching function status
SP1 … SP6 - Switching
function 1 … 6