84 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
Transmit: SCx [,a,b,c.ccEscc,d.ddEsdd] <CR>[<LF>]
Description
x Controlled gauge, x =
1 –> Gauge 1
2 –> Gauge 2
3 –> Gauge 3
a Gauge activation, a =
0 –> Manual (default)
1 –> Hot start
3 –> Via measurement channel 1
4 –> Via measurement channel 2
5 –> Via measurement channel 3
b Gauge deactivation, b =
0 –> Manual (default)
1 –> Self control
3 –> Via measurement channel 1
4 –> Via measurement channel 2
5 –> Via measurement channel 3
c.ccEscc ON threshold (s = sign)
d.ddEsdd OFF threshold (s = sign)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c.ccEscc,d.ddEsdd <CR><LF>
Description
a Gauge activation
b Gauge deactivation
c.ccEscc ON threshold (s = sign)
d.ddEsdd OFF threshold (s = sign)
5.7 Gauge Control
SC1, SC2, SC3 - Gauge 1, 2 or
3 control