8 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
2 Technical Data
Voltage 100 … 240 V (ac) ±10%
Frequency 50 … 60 Hz
Power consumption
VGC501
VGC502
VGC503
≤45 W
≤65 W
≤90 W
Overvoltage category II
Protection class 1
Connection European appliance connector IEC 320 C14
Temperature
Storage
Operation
–20 … +60 °C
+ 5 … +50 °C
Relative humidity ≤80% up to +31 °C,
decreasing to 50% at +40 °C
Use indoors only
max. altitude 2000 m NN
Pollution degree II
Degree of protection IP30
Number
VGC501
VGC502
VGC503
1
2
3
Gauge connections per channel
RJ45 (FCC68), 8-pin (→ 20)
D-Sub, 15-pin, female (→ 20)
(connected in parallel)
Compatible gauges
Pirani PSG400, PSG400-S, PSG100-S, PSG101-S,
PSG500, PSG500-S, PSG502-S, PSG510-S,
PSG512-S, PSG550, PSG552, PSG554
Pirani / Capacitance PCG400, PCG400-S, PCG550, PCG552,
PCG554
Cold cathode PEG100, MAG500, MAG504
Cold cathode / Pirani MPG400, MPG401, MPG500, MPG504
Hot ionization / Pirani BPG400, BPG402, HPG400
Capacitance CDG020D, CDG025, CDG025D,
CDG025D-X3, CDG045, CDG045-H,
CDG045D, CDG045D2, CDG045Dhs,
CDG100, CDG100D, CDG100D2,
CDG100Dhs, CDG160D, CDG200D
Hot ionization / Capacitance /
Pirani
BCG450
Voltage +24 V (dc) ±5%
Ripple <±1%
Current 0 … 1 A (per channel)
Power 25 W (per channel)
Fuse protection 1.5 A (per channel) with PTC element, self-
resetting after turning the unit off or disconnect-
ing the gauge. The supply conforms to the
grounded protective extra low voltage require-
ments.
Mains specifications
Ambience
Gauge connections
Gauge supply