92 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
The group is only available when a USB memory stick formatted for the
the FAT file system (FAT32) is plugged in. Use a max. 32 GB memory
stick.
Transmit: SCM [,a,bb] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a <CR><LF>
Description
a Setup parameters, a =
0 –> Saving completed (read only)
1 –> CSV file is being saved (read only)
2 –> Loading all parameters from the USB memory stick
onto the VGC50x
3 –> Formatting USB memory stick (FAT32)
4 –> Deleting parameter files (ending CSV) from the USB
memory stick
bb Number in the file name (0 … 99)
(For service personnel)
ADC corresponds to the TAD command
Transmit: CPT [,a] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a <CR><LF>
Description
a a =
0 –> INFICON gauges (standard)
1 –> OLV transmitter
DIS corresponds to the TDI command
EEP corresponds to the TEE command
EPR corresponds to the TEP command
5.10 Parameter Transfer
SCM - Save / load parameters
(USB)
5.11 Test Parameters
ADC - A/D converter test
CPT - Compatibility
DIS - Display test
EEP - EEPROM test
EPR - FLASH test