tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 93
Transmit: HDW <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a.a <CR><LF>
Description
a.a Hardware version, e.g. 1.0
IOT corresponds to the TIO command
Transmit: LOC [,a] <CR>[<LF>]
Description
a Keylock, a =
0 –> Off (default)
1 –> On
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a <CR><LF>
Description
a Keylock status
Transmit: MAC <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: aa-aa-aa-aa-aa-aa <CR><LF>
Description
aa-aa-aa-aa-aa-aa Ethernet MAC address of the unit:
00-A0-41-0A-00-00 … 00-A0-41-0B-FF-FF
Transmit: PNR <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a.aa <CR><LF>
Description
a.aa Firmware version, e.g. 1.00
Transmit: RHR <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a <CR><LF>
Description
a Run (operating) hours, e.g. 24 [hours]
HDW - Hardware version
IOT - I/O test
LOC - Keylock
MAC - Ethernet MAC address
PNR - Firmware version
RHR - Operating hours