90 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
Transmit: PRE [,a] <CR>[<LF>]
Description
a Pirani range extension, a =
0 –> Disabled (default)
1 –> Enabled
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a <CR><LF>
Description
a Pirani range extension status
PCG and PSG gauges only, measurement range up to 5×10
-5
mbar.
Transmit: SAV [,a] <CR>[<LF>]
Description
a Save parameters to EEPROM, a =
0 –> Save default parameters (default)
1 –> Save user parameters
Receive: <ACK><CR><LF>
Transmit: UNI [,a] <CR>[<LF>]
Description
a Pressure unit, a =
0 –> mbar/bar
1 –> Torr
2 –> Pascal
3 –> Micron
4 –> hPascal (default)
5 –> Volt
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a <CR><LF>
Description
a Pressure unit
PRE - Pirani range extension
SAV - Save parameters
(EEPROM)
UNI - Pressure unit