52 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
Value
AO-MODE I
AO-MODE LOG I Logarithmic representation of the entire
measuring range (factory setting).
PSG: p = 10^[U/(10 / 7) - 4]
PCG: p = 10^[U/(10 / 7) - 4]
PEG/MAG: p = 10^[U/(10 / 7) - 9]
MPG: p = 10^[U/(10 / 12) - 9]
CDG: p = 10^[U/(10 / 4) - 4] × FS
BPG: p = 10^[U/(10 / 12) - 9]
BCG: p = 10^[U/(10 / 12) - 9]
HPG: p = 10^[U/(10 / 9) - 6]
AO-MODE LOG A I Logarithmic representation of the entire
measuring range (compatible to VGC012,
VGC023, VGC032).
PSG: p = 10^[U/(10/6) - 3]
PCG: p = 10^[U/(10/7) - 4]
PEG/MAG: p = 10^[U/(9/7) - 9 - 7/9]
MPG: p = 10^[U/(10/11) - 8]
CDG: p = 10^[U/(10/4) - 4] × FS
BPG400: p = 10^[(U - 7.75) / 0.75]
BPG402: p = 10^[U - 8]
BCG: p = 10^[(U - 7.75) / 0.75]
HPG: p = 10^[U/(10/9) - 6]
AO-MODE LOG -6 I Logarithmic representation of a part of the
measuring range (2.5 V/decade).
p = 10^[U/(10/4) - 10]
AO-MODE LOG -3 I Logarithmic representation of a part of the
measuring range (2.5 V/decade).
p = 10^[U/(10/4) - 7]
AO-MODE LOG +0 I Logarithmic representation of a part of the
measuring range (2.5 V/decade).
p = 10^[U/(10/4) - 4]
AO-MODE LOG +3 I Logarithmic representation of a part of the
measuring range (2.5 V/decade).
p = 10^[U/(10/4) - 1]
AO-MODE LOG C1 I Logarithmic representation matched to the
following sensor combination:
PSG/PCG on channel 1
PEG on channel 2
p = 10^[U/(10/12) - 9]
AO-MODE LOG C2 I Logarithmic representation matched to the
following sensor combination:
CDG on channel 1
CDG on channel 2
This characteristic curve is only useful if the
sensors have different measuring ranges.
The total measuring range of the sensor
combination is represented logarithmically
in the range 0…10 V.