EasyManua.ls Logo

Inficon VGC501 - Page 73

Inficon VGC501
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 73
Transmit: PRX <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,sx.xxxxEsxx,b,sy.yyyyEsyy,c,sz.zzzzEszz <CR><LF>
Description
a Status gauge 1, a =
0 –> Measurement data okay
1 –> Underrange
2 –> Overrange
3 –> Sensor error
4 –> Sensor off (PEG, MAG)
5 –> No Sensor
6 –> Identification error
7 –> Error BPG, HPG, BCG
sx.xxxxEsxx Measurement value gauge 1
1)
[in current
pressure unit] (s = sign)
b Status gauge 2
sy.yyyyEsyy Measurement value gauge 2
1)
[in current
pressure unit] (s = sign)
c Status gauge 3
sz.zzzzEszz Measurement value gauge 3
1)
[in current
pressure unit] (s = sign)
1)
Values always in exponential format.
For logarithmic gauges, the 3
rd
and 4
th
decimal are always 0.
PRX - Measurement data
gauge 1, 2 and 3

Related product manuals