114 M10-17410-EN
Table 10. Lubrication Requirements for Ballnuts
Recommended Lubricant Interval Instron
®
Part Number
Lithium based water resisting grease -
NLGI Class 2
(DIN51825 class K2k LS2)
Equivalents:
Lubriplate: Lubriplate 1200-2
Mobil: Mobilux® 2
BP: Energrease® LS2
Shell: Alvania® R2
1000
1
hours or 36
months of
operating time
2
105-1-1057
(small can of lubricant)
1. If the system is used more than 100 hours per month, inspect the lubricated area periodically and
increase lubrication frequency as required.
2. Operating time is defined as time of crosshead movement, not the length of time the system has been
powered on.