Manual de instrucciones del AT235 - ES Página 46
Tabla 1: Intervalos de intensidad y frecuencia
AT235 Maximums IMP
TDH39
CIR33
EARtone 3A / IP30
IPSI
DD45
Center
Reading
Reading
Reading
Reading
Reading
Freq.
Tone
NB
Tone
NB
Tone
NB
Tone
NB
Tone
NB
[Hz]
[dB HL]
[dB HL]
[dB HL]
[dB HL]
[dB HL]
[dB HL]
[dB HL]
[dB HL]
[dB HL]
[dB HL]
125
80
65
90
70
100
85
70
60
75
60
250
100
85
100
85
110
100
85
75
100
85
500
120
100
110
100
115
105
100
85
115
100
750
120
105
110
105
120
110
100
85
120
105
1000
120
105
115
105
120
110
105
90
120
105
1500
120
105
115
105
120
110
110
90
115
100
2000
120
105
115
105
120
110
105
90
115
100
3000
120
105
115
105
120
110
95
90
120
105
4000
120
105
110
100
120
105
100
85
115
105
6000
120
100
95
95
105
100
85
80
110
90
8000
105
95
75
80
90
85
80
75
105
95
10000
WB
-
115
-
115
-
115
-
95
-
120
LP
-
120
-
115
-
120
-
100
-
120
HP
-
115
-
115
-
120
-
95
-
120

Table of Contents

Other manuals for Interacoustics AT235

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics AT235 and is the answer not in the manual?

Interacoustics AT235 Specifications

General IconGeneral
BrandInteracoustics
ModelAT235
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals