AT235 Instructions for Use - EN Page 15
3.3 Probe Status
The probe status is indicated by the color of the light on the control panel, standard probe system and the
clinical probe system. Below, the colors and their meanings are explained:
Color
Control panel
Standard probe
Clinical probe
Status
Red
Right ear is selected. Probe is out of
the ear.
Blue
Left ear is selected. Probe is out of
the ear.
Green
Probe is in the ear and a seal is
maintained.
Yellow
Probe is in the ear and is blocked,
leaking or too noisy.
White
The probe has just been attached.
Probe status is unknown. If the probe
light stays white in any other situation
the AT235 may need to be switched
off and back on again to regain
proper probe status.
Blinking
AT235 is pausing and/or an
interaction is expected. The AT235
will for example remain blinking green
if the protocol has finished testing and
the probe is still in the ear. The user
may pause the AT235 before
inserting the probe resulting in blue or
red blinking.
No light
AT235 is not monitoring the probe
status.

Table of Contents

Other manuals for Interacoustics AT235

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics AT235 and is the answer not in the manual?

Interacoustics AT235 Specifications

General IconGeneral
BrandInteracoustics
ModelAT235
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals