AT235 Instructions for Use - EN Page 45
5 Specifications
5.1 AT235 Technical Specifications
Medical CE-mark: The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II
of the Medical Device Directive 93/42/EEC.
Approval of the quality system is made by TÜV identification no0123
Standards:
Safety:
IEC 60601-1, Class I, Type B applied parts
EMC:
IEC 60601-1-2
Impedance:
IEC 60645-5/ANSI S3.39, Type 1
Audiometer:
IEC60645-1/ANSI S3.6, Type 4
Operation
environment:
Temperature:
15 35 °C
Relative Humidity:
30 90%
Ambient Pressure:
98kPa 104kPa
Warm-up Time:
1 minute
Transport & Storage: Storage Temperature:
Transport Temperature:
Rel. Humidity:
0°C 50°C
-20 50 °C
10 95%
Internal Battery
CR2032 3V, 230mAh, Li. Not serviceable by user.
Impedance Measuring System
Probe tone: Frequency:
Level:
226 Hz, 678 Hz, 800 Hz, 1000 Hz; pure tones; ±1%
85 dB SPL (≈ 69 dB HL) ±1.5 dB
Air pressure: Control:
Indicator:
Range:
Pressure limitation:
Automatic.
Measured value is displayed on the graphical display.
-600 to +400 daPa. ±5%
-750 daPa and +550 daPa.
Compliance: Range: 0.1 to 8.0 ml at 226 Hz probe tone (Ear volume: 0.1 to
8.0 ml) and 0.1 to 15 mmho at 678, 800 and 1000 Hz
probe tone. All ±5%
Test types: Tympanometry Automatic, where the start and stop pressure can be
user-programmed in the setup function.
Manual control of all functions.

Table of Contents

Other manuals for Interacoustics AT235

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics AT235 and is the answer not in the manual?

Interacoustics AT235 Specifications

General IconGeneral
BrandInteracoustics
ModelAT235
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals