AT235 Instructions for Use - EN Page 51
Terminal latency = 23 ms (5 ms)
Fall time = 44 ms (5 ms)
Overshoot = max. 1%
Undershoot = max 1%
Values between -600 to +400 daPa can be selected in
-750 daPa and +550 daPa, 50 daPa
Reflex Calibration Standards and Spectral Properties:
Specifications for stimulus and audiometer signals are made to follow IEC 60645-5
ISO 389-1 for TDH39 and ISO 389-2 for CIR 55.
As “Broad band noise” specified in IEC 60645-5, but
with 500 Hz as lower cut-off frequency.
Uniform from 500 Hz to 1600 Hz, 5 dB re. 1000 Hz
level
Uniform from 1600 Hz to 10KHz, 5 dB re. 1000 Hz
level
As “Broad band noise” specified in IEC 60645-5, but
with 500 Hz as lower cut-off frequency.
Uniform from 500 Hz to 1600 Hz, 10 dB re. 1000 Hz
level
Uniform from 1600 Hz to 4000 Hz, 10 dB re. 1000 Hz
level
The actual sound pressure level at the eardrum will
depend on the volume of the ear.
The risk of artifacts at higher stimulus levels in reflex measurements are minor and will not activate the