13
Example, with the FlexiPump
®
:
FlexiPump
®
Pro :
How to read the diagram :
For a given tubing diameter, the curve indicates the guaranteed precision according to the volume of a
dose. Example : in this figure, for a tubing with inner diameter of 4,8 mm (red curve) and a volume of 9
mL, the guaranteed precision is inferior to 4% (follow the dots in red).
Tube diameters
Volume
Ø
8
Ø6,
Ø4
,
Ø3,
Ø1,
Precision