28
5 Maintenance
FlexiPump
®
requires little maintenance.
- Daily :
Autoclave the tubing set (when the FlexiPump
®
is used in a sterile environment).
The tubing supports are fully autoclavable.
- Change the silicone tubing :
In the event of leakage, hardening or sticking of the tubing.
- Every 20 autoclavaging cycles :
Filter change-to-air (only for operation in a sterile environment).
Note : INTERSCIENCE can provide complete dispensing assembly.
Note : To clean the top shell, do not use bleach because it damages the stainless steel. We recommend
the use of INTERSCIENCE Clinet
®
101 disinfectant wipes.