General Care Instructions
NOTE: Do
not disassemble the InterX 5002. Dangerous voltages could be
present. The InterX 5002 does not contain any user-serviceable components.
If the device needs repair or service, contact InterX Technologies, your
InterX distributor or an authorized service representative.
Do not expose any part of the InterX 5002 to chemical solvents or harsh
cleaning fluids. Do not sterilize or immerse the InterX 5002 in any fluid.
Storage and Cleaning
Remove the batteries when storing the InterX 5002 for more than one month.
Storage in extreme heat or cold will deplete battery life.
For protection purposes, use the black holster to transport the InterX 5002.
When not in use the InterX 5002 should be turned off.
Clean the device and accessory electrodes with the main power off. Between
patient treatments, thoroughly clean the main electrode, accessory electrodes,
and surrounding device area with 70% isopropyl alcohol wipes. Never
spray cleaners directly on the device.
The InterX 5002 is a non-critical patient contact device indicated only for
contact between the electrodes and intact skin.
CAUTION: Do not use cleaning products that contain ethyl alcohol and/or
ammonium chlorides. These chemicals may cause cracking of the plastic.
The only approved cleaning agent is isopropyl alcohol that is less than or
equal to 90% by volume.
Using unapproved cleaning agents will void the manufacturer’s warranty.