SECTION 3—TROUBLESHOOTING
Part No 1114836 19 IVC
Bed Series
Electrical Inspection and Maintenance
Inspect all electrical bed components for damage or excessive wear (i.e. cracked or
broken housings, or worn components).
Check pendant, power and motor cords for chafing, cuts or excessive wear.
Make sure all plugs are fully attached and free of damage.
Make sure cable lock on junction box is properly positioned and locked
Check all functions:
Head raises and lowers properly.
Foot raises and lowers properly.
Bed Ends raise and lower properly.
Pass/Fail
PASS: Bed and all components have passed all of the above inspection criteria and the
bed is ready for placement into a client’s home.
FAIL: Bed, or any component, has failed any of the above inspection criteria. Tag the
bed or component with a complete description of the failure(s) and have the bed
serviced.