Troubleshooting
7Troubleshooting
7.1Troubleshooting
SYMPTOMSFAULTSSOLUTION
Patientliftfeelsloose.
Mast/Basejointloose.
RefertotheAssemblingtheMasttothe
Basesectioninthemanual.
Casters/Brakesnoisyorstiff.
Lint,dirt,debrisorthelikeinbearings.RefertoReplacingFrontCastersand
ReplacingRearCasterssectionsinthe
manual.
Noisyordrysoundfrompivots.Needslubrication.RefertoLubricatingtheLiftsectionin
themanual.
Pendantoractuatorconnectorloose.Connectpendantoractuatorconnector .
Ensureconnectorsareseatedproperly
andfullyconnected.
Batterylow.Chargebatteries.RefertoChargingthe
Batterysectioninthemanual.
REDemergencystopbuttonpressedIN.RotateREDemergencystopbutton
CLOCKWISEuntilitpopsout.
Batterynotconnectedproperlyto
controlbox.
Reconnectthebatterytothecontrol
box.RefertoChargingtheBattery
sectioninthemanual.
Theconnectingterminalsaredamaged.Replacethebatterypack.Referto
ChargingtheBatterysectioninthe
manual.
Electricactuatorfailstoliftorlegsfailto
openwhenbuttonispressed.