Invacare®Jasmine™
Guidanceandmanufacturer´sdeclaration–electromagneticimmunity
Thepatientliftisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthepatient
liftshouldassurethatitisusedinsuchanenvironment.
Immunitytest
IEC60601testlevelCompliancelevelElectromagneticenvironment–guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Electrostatic
transient/burst
IEC61000-4-4
±2kVforpowersupply
lines
±1kVforinput/output
lines
±2kVforpowersupply
lines
±1kVforinput/output
lines
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Surge
IEC61000-4-5
±1kVline(s)toline(s)±1kVline(s)toline(s)
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Productisdouble-insulated.Therearenoother
possibleconnectionstoearth
Voltagedips,short
interruptionsand
voltagevariations
onpowersupply
inputlines
IEC61000-4-11
<5%U
T
(>95%dipinU
T
)
for0,5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipinU
T
)
for5sec
<5%U
T
(>95%dipinU
T
)
for0,5cycle
40%U
T
(60%dipinU
T
)for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipinU
T
)
for5sec
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuserof
thepatientliftrequirescontinuedoperationduring
powermainsinterruptions,itisrecommended
thatthepatientliftbepoweredfroman
un-interruptiblepowersupplyorabattery.
U
T
isthea.c.mainsvoltagepriortoapplication
ofthetestlevel.
641195053-B