SECTION 8—TROUBLESHOOTING GUIDE
Part No 1160877 27 Perfecto
2
™Series
SECTION 8—TROUBLESHOOTING
GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
Alarm: Main Power Loss:
Short beeps, long
pause
1. Power cord not
plugged in.
1. Insert plug into outlet.
Concentrator not
operating,
power switch On.
2. No power at outlet. 2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
Beep......Beep...... 3. Tripped circuit
breaker.
3. Push/reset circuit breaker.
If problem recurs, call Service
Provider.
Alarm:
Continuous
Concentrator not
operating, Power
Switch On.
Beeeeeeeep....
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required.
1a. Remove and clean cabinet
filters.
1b. Move oxygen
concentrator at least three
inches away from walls,
draperies or furniture.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.

Other manuals for Invacare Perfecto2 IRC5PO2AW

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 IRC5PO2AW and is the answer not in the manual?

Invacare Perfecto2 IRC5PO2AW Specifications

General IconGeneral
Flow Rate0.5 to 5 LPM
Warranty3 years limited
Operating Environment10°C to 35°C; up to 95% relative humidity
Power Requirements120 VAC, 60 Hz
Sound Level≤45 dB(A)
Dimensions23 in (H) x 13 in (W) x 12 in (D) (58.4 cm (H) x 33 cm (W) x 30.5 cm (D))

Related product manuals