SECTION 5—TYPICAL PRODUCT PARAMETERS
Perfecto
2
™Series 16 Part No 1160877
IRC5PO2AW & IRC5PO2VAW INDICATORS
IRC5PAW INDICATORS
LABEL
SYMBOL
O
2
PURITY INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O
2
over 85% (± 2%)
GREEN Indicator Light
O
2
Between 73% (± 3%)
to 85% (±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73% (±3%)
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor
Shutdown
LABEL
SYMBOL
STATUS INDICATOR LIGHTS
(LED)
SYSTEM OKAY GREEN Indicator Light
SYSTEM FAILURE RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician.
O
2
I/O

Other manuals for Invacare Perfecto2 IRC5PO2AW

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 IRC5PO2AW and is the answer not in the manual?

Invacare Perfecto2 IRC5PO2AW Specifications

General IconGeneral
Flow Rate0.5 to 5 LPM
Warranty3 years limited
Operating Environment10°C to 35°C; up to 95% relative humidity
Power Requirements120 VAC, 60 Hz
Sound Level≤45 dB(A)
Dimensions23 in (H) x 13 in (W) x 12 in (D) (58.4 cm (H) x 33 cm (W) x 30.5 cm (D))